TOSHIBA E7239X X-ray tube Unit
TOSHIBA E7239X X-ray Tube Unit
The TOSHIBA E7239X X-ray tube unit is a rotating-anode diagnostic X-ray source designed for compatible medical radiography equipment. Manufactured by TOSHIBA under catalog number E7239X, this dual-focus tube unit supports routine diagnostic imaging with selectable focal spots for balancing image detail and tube loading.
Purpose and Key Features
- Rotating-anode construction for diagnostic X-ray generation
- Dual focal spots for different radiographic exposure requirements
- Maximum tube voltage of 125 kV
- Integrated tube-unit design intended for installation in approved X-ray systems
- Suitable as a replacement only where E7239X is specified by the equipment manufacturer
Technical Specifications
- Manufacturer: TOSHIBA
- Model and catalog number: E7239X
- Product type: Rotating-anode diagnostic X-ray tube unit
- Maximum X-ray tube voltage: 125 kV
- Nominal focal spots: 1.0 mm and 2.0 mm
- Target angle: 16 degrees
- Focus configuration: Dual focus
Compatible Medical Equipment
The TOSHIBA E7239X is compatible with diagnostic radiography systems whose official parts list, service documentation, or installed tube-unit nameplate specifies E7239X. It is a direct model match for equipment originally configured for the TOSHIBA E7239X tube unit. Compatibility must not be determined from tube voltage or focal-spot size alone.
Before ordering, verify the complete model and catalog number, generator requirements, electrical connections, stator compatibility, mounting arrangement, and available installation space against the medical device service manual. Substitution in a system specifying another X-ray tube model requires written approval from the equipment manufacturer or an authorized technical service organization.
Installation and Usage Instructions
- Confirm that the medical X-ray system documentation explicitly specifies the TOSHIBA E7239X.
- Have the tube unit installed only by a qualified medical X-ray service engineer with the equipment disconnected and secured according to applicable electrical-safety procedures.
- Follow the system manufacturer's instructions for mounting, grounding, high-voltage connection, stator connection, and generator configuration.
- After installation, complete all prescribed electrical-safety tests, tube seasoning, warm-up, alignment, calibration, and image-quality checks before clinical use.
- Operate the tube only within the approved exposure loading charts, duty cycle, and cooling limits provided for the X-ray system and tube unit.
- Stop operation and request technical inspection if oil leakage, electrical arcing, unusual noise, excessive heat, unstable output, or physical damage is detected.
The tube housing is not intended to be opened or repaired by clinical users. Installation, testing, and servicing must comply with the medical device manufacturer's procedures and local radiation-safety regulations.
Order from Partsformed
Partsformed supplies high-quality medical imaging products at competitive prices and offers worldwide shipping. When requesting the TOSHIBA E7239X X-ray tube unit, provide the X-ray system model, serial number, installed tube label details, and required delivery destination to support accurate identification and efficient order processing.