Inspire Innovations international limited.
Type the part number or keyword in the search box. If you do not know the catalog number, you can find the product by category or by manufacturer. To find a product by device model, first select the manufacturer.
OXYGEN SOLENOID ASSEMBLY by Puritan Bennett - Covidien
Oxygen Solenoid Assembly G-062023-00 by Puritan Bennett–Covidien
Product Purpose and Features
The Oxygen Solenoid Assembly G-062023-00, manufactured by Puritan Bennett, is a replacement service component for compatible Puritan Bennett–Covidien medical equipment. This electromechanical assembly is designed to actuate an oxygen-flow path within the host device under its electronic control.
The assembly is intended for professional equipment maintenance rather than standalone use. Matching the complete catalog number is essential because electrical, pneumatic, and mechanical configurations can differ between device revisions.
- Solenoid-operated component for an oxygen circuit
- Puritan Bennett catalog number for precise part identification
- Assembly format suitable for authorized repair and preventive maintenance
- Designed for integration with the host device’s electrical and pneumatic systems
Technical Specifications
- Product name: Oxygen Solenoid Assembly
- Catalog number: G-062023-00
- Manufacturer: Puritan Bennett
- Brand association: Puritan Bennett–Covidien
- Product type: Oxygen solenoid assembly
- Operating principle: Electromagnetic solenoid actuation
- Gas-circuit designation: Oxygen
- Application: Internal replacement and service component
- Installation: Professional installation inside compatible medical equipment
Compatible Medical Devices and Models
The Puritan Bennett G-062023-00 oxygen solenoid assembly is compatible only with Puritan Bennett–Covidien medical devices and specific equipment revisions whose service documentation or parts list identifies catalog number G-062023-00. Compatibility should be confirmed using the device model, serial-number range, installed component label, and current manufacturer service documentation. Brand similarity or physical appearance alone does not establish compatibility.
Installation and Usage Instructions
- Confirm that G-062023-00 is the specified replacement part for the exact device model and revision.
- Remove the medical device from clinical service and disconnect electrical power and all gas supplies.
- Depressurize the pneumatic system according to the equipment service manual.
- Document the original orientation and the positions of electrical connectors, tubing, fittings, and mounting hardware.
- Remove the existing assembly using the manufacturer-approved service procedure.
- Install the replacement oxygen solenoid assembly in the specified orientation without twisting tubing or damaging connectors and seals.
- Reconnect all electrical and pneumatic interfaces and secure the assembly with the specified hardware.
- Perform the required oxygen-circuit leak test, functional test, alarm verification, and any prescribed calibration before returning the device to service.
Installation, testing, and servicing must be carried out by qualified biomedical or authorized technical personnel. Follow the applicable device service manual and healthcare-facility procedures for oxygen-system safety.