Inspire Innovations international limited.
Type the part number or keyword in the search box. If you do not know the catalog number, you can find the product by category or by manufacturer. To find a product by device model, first select the manufacturer.
Draeger (Drager) 8103061 DETECTOR TUBE, WATER VAPOR
Draeger (Drager) 8103061 Detector Tube, Water Vapor
Product Description
The Draeger (Drager) 8103061 Detector Tube, Water Vapor, designated Water Vapour 0.1/a, is a direct-reading colorimetric tube for measuring low concentrations of water vapor in compressed air and compressed gases. Manufactured by Drager, this single-use glass detector tube provides an on-site indication without requiring laboratory analysis.
During sampling, the reagent changes from yellow to reddish brown. The length of the discoloration is evaluated against the printed tube scale to determine the water vapor concentration. The Draeger 8103061 tube is suitable for compressed-gas quality checks performed within its specified measuring range and operating conditions.
Technical Specifications
- Manufacturer: Draeger (Drager)
- Catalog number: 8103061
- Drager order format: 81 03 061
- Product designation: Water Vapour 0.1/a
- Product type: Direct-reading colorimetric detector tube
- Measured substance: Water vapor, H2O
- Measuring range: 0.1–1.0 mg/L
- Required pump strokes: 10
- Measurement time: Approximately 10 minutes
- Standard deviation: ±15–20%
- Color change: Yellow to reddish brown
- Operating temperature: 0–40 °C
- Sample type: Depressurized compressed air or compressed gas
- Use format: Single-use tube
Compatible Equipment and Medical Applications
The Draeger 8103061 water vapor tube is used with compatible Drager tube-sampling equipment, including the Drager accuro manual tube pump and the Drager X-act 5000 automatic tube pump, when configured and operated according to the applicable instructions.
This detector tube is not installed inside a ventilator, anesthesia workstation, patient monitor, or other medical device. Its compatibility applies to gas-sampling equipment rather than to a specific medical device model. In healthcare environments, it may support testing of appropriately collected and depressurized compressed-gas samples, including air-quality verification programs, subject to the facility’s approved procedures and applicable standards.
Usage Instructions
- Review the current instructions for the detector tube and sampling pump before use. Confirm that the expected water vapor concentration and sample temperature are within the specified limits.
- Obtain a representative, depressurized sample. Never connect the glass tube or tube pump directly to a pressurized gas outlet without an approved pressure-reduction and sampling arrangement.
- Open both sealed ends of the tube with an approved Drager tube opener while protecting the tube from breakage and sharp glass fragments.
- Insert the tube securely into the compatible pump with the arrow pointing toward the pump.
- Draw exactly 10 complete pump strokes through the tube. Allow each stroke to finish before beginning the next one.
- After sampling, read the end of the reddish-brown discoloration directly against the printed scale and record the result in mg/L. Apply only the corrections specified in the current Drager instructions for the actual sampling conditions.
- Dispose of the used tube as glass and chemical waste in accordance with local regulations and workplace procedures. Do not reuse the tube.
Ordering from Partsformed
Partsformed supplies the Draeger (Drager) 8103061 Detector Tube, Water Vapor for professional gas-testing and compressed-air quality applications. Customers benefit from worldwide shipping, competitive prices, and high-quality products. Verify the catalog number, measuring range, sampling equipment, and intended application before ordering.