Hamilton Medical AG 281637 Proximal flow sensor adult/pediatric 10/pk
Hamilton Medical AG 281637 Proximal flow sensor adult/pediatric 10/pk
Flow range 0 to 180 l/min (short term 260 l/min)
Dead space < 11 ml (functional)
Resistance < 0.027 cmH2O/(l/min) over the complete flow range (-180 l/min to 180 l/min)
Weight Approx. 14 g without tubes, 68 g with 188 cm tubes, 89 g with 260 cm tubes, 110 g with 330 cm tubes
Material Complies with ISO 10993-1:2009 (biocompatibility) and Directive 2011/65/EU (RoHS).
Does not contain natural rubber latex, DEHP, or BPA.
Accuracy For calibrated flow sensor volume-related parameters:
HAMILTON-C1/T1/MR1/C2/C3/C6 ventilators: ± 10% or ± 10 ml (whichever is greater)
HAMILTON-G5/S1 ventilators: ± 15% or ± 20 ml (whichever is greater)
Temperature Operating: -20°C to 50°C (-4°F to 122°F)
Storage: -20°C to 60°C (-4°F to 140°F)
Operating time 28 days or depending on hospital infection control protocols, patient secretions, and nebulization of
medication
Accessory Tube clip, calibration adapter ID22 / OD15
Identification Individual serial number